Operator's Manual
14
TF-DSQ-OM-E RevC
Probe Calibration
Probe Calibration Menu
Accessing the Probe Calibration Menu
The probe calibration menu can be accessed from the Vertical drop-down menu or from the
Channel (sometimes referred to as Vertical Adjust) dialog:
Probe Calibration Menu Description
Figure 4-5. The Probe Calibration dialog contains information and controls.
The information on the probe calibration dialog is organized so each row represents the
information for a given channel and each column represents the calibration information or
control for that channel. For each channel, the information and controls provided include:
The channel number in the colored button icon and the probe type that is installed
A Full Calibration button, which starts the calibration
DC correction information including both gain and offset correction
The skew correction
A Clear button